Manometer Basics

Manometer Basics

By: Rick D'Angelo,Dave Thomas,


One of the earliest pressure measuring instruments is still in wide use today because of its inherent accuracy and simplicity of operation. It's the U-tube manometer, which is a U-shaped glass tube partially filled with liquid. This manometer has no moving parts and requires no calibration. Manometry measurements are functions of gravity and the liquid's density, both physical properties that make the U-tube manometer a NIST standard for accuracy.
photo


Manometers are both pressure measurement instruments and calibration standards. They range from simple U-tubes and wells filled with liquid to portable digital instruments with a computer interface.
As shown in Figure 1, with each leg of a U-tube manometer exposed to the atmosphere, the height of liquid in the columns is equal. Using this point as a reference and connecting each leg to an unknown pressure, the difference in column heights indicates the difference in pressures (see Figure 2).
figure


Figure 1. With both legs of a U-tube manometer open to the atmosphere or subjected to the same pressure, the liquid maintains the same level in each leg, establishing a zero reference.
figure


Figure 2. With a greater pressure applied to the left side of a U-tube manometer, the liquid lowers in the left leg and rises in the right leg. The liquid moves until the unit weight of the liquid, as indicated by h, exactly balances the pressure.
The fundamental relationship for pressure expressed by a liquid column is:
Δp = P2-P1 = ρgh(1)
where:
Δp = differential pressure
P1= pressure at the low-pressure connection
P2= pressure at the high-pressure connection
ρ= density of the indicating fluid (at a specific temperature)
g= acceleration of gravity (at a specific latitude and elevation)
h= difference in column heights
The resulting pressure is the difference between forces exerted per unit of surface area of the liquid columns, with pounds per square inch (psi) or newtons per square meter (pascals) as the units. The manometer is so often used to measure pressure that the difference in column heights is also a common unit. This is expressed in inches or centimeters of water or mercury at a specific temperature, which can be changed to standard units of pressure with a conversion table.
All pressure measurements are differential. The reference can be zero absolute pressure (a total vacuum), atmospheric pressure (the barometric pressure), or another pressure. With one leg of a manometer open to the atmosphere (see Figure 3A), the measured pressure is that which exceeds atmospheric pressure, which at sea level is 14.7 psi, 101.3 kPa, or 76 cmHg.
figure


Figure 3. Gauge pressure is a measurement relative to atmospheric pressure and it varies with the barometric reading. A gauge pressure measurement is positive when the unknown pressure exceeds atmospheric pressure (A), and is negative when the unknown pressure is less than atmospheric pressure (B).
This measurement is called gauge pressure, and the relationship for a positive pressure is expressed by:
absolute pressure = atmospheric pressure + positive gauge pressure(2)
For a negative pressure (vacuum) measurement (see Figure 3B), the column heights reverse and the relationship is expressed by:
absolute pressure = atmospheric pressure + negative gauge pressure(3)
These pressure relationships are shown in Figure 4.
figure


Figure 4. A graphical representation of positive and negative gauge pressure shows the differential aspect of all pressure measurements, where gauge pressure is the difference between absolute pressure and atmospheric pressure.
figure


Figure 5. In a sealed-tube manometer, the pressure reference is a vacuum, or zero absolute pressure. The most common form of a sealed-tube manometer is the conventional mercury barometer used to measure atmospheric pressure.
A manometer can be designed to directly measure absolute pressure. The manometer in Figure 5 measures the pressure compared to zero absolute pressure in a sealed leg above a mercury column. The most common form of this manometer is the conventional mercury barometer used to measure atmospheric pressure. With just one connection, this configuration can measure pressures above and below atmospheric pressure.
Variations on the U-Tube Manometer


The differential pressure is always the difference in column heights, regardless of the size or shape of the tubes. As shown in Figure 6A, the legs of both manometers are open to the atmosphere and the indicating fluids are at the same level. Connecting the same pressure to the left leg of each manometer causes its level to lower. Because of the variation in volume in the manometer legs, the fluid in each column moves a different distance. However, the difference between the fluid levels in both manometers is identical (see Figure 6B).

figure


Figure 6. The pressure reading is always the difference between fluid heights, regardless of the tube sizes. With both manometer legs open to the atmosphere, the fluid levels are the same (A). With an equal positive pressure applied to one leg of each manometer, the fluid levels differ, but the distance between the fluid heights is the same.
figure


Figure 7. In a well-type manometer, the cross-sectional area of one leg (the well) is much larger than the other leg. When pressure is applied to the well, the fluid lowers only slightly compared to the fluid rise in the other leg.
Carrying this variation in tube sizes further is the well-type (or reservoir) manometer (see Figure 7). As pressure is applied to the well, the level falls slightly as compared to the level rise in the column. By compensating the column's scale graduations to correct for the well drop, it is possible to make a direct reading of differential pressure. There are connection guidelines placed on well-type manometers, compared to the U-tube style:
  • Connect pressures higher than atmospheric to the well; connect pressures lower than atmospheric to the tube.
  • For differential measurements, connect the higher pressure to the well.
  • For raised-well manometers, the well connection can be used for gauge and vacuum measurements.
A variation of the well-type manometer is the inclined-tube (or draft gauge) manometer in Figure 8. With an inclined indicating tube, 1 in. of a vertical rise is stretched over several inches of scale length. The inclined-tube manometer has better sensitivity and resolution for low pressures.
figure


Figure 8. Low pressure and low differentials are better handled with an inclined-tube manometer, where 1 in. of vertical liquid height can be stretched to 12 in. of scale length.
Indicating Fluids
Liquid manometers measure differential pressure by balancing the weight of a liquid between two pressures. Light liquids such as water can measure small pressure differences; mercury or other heavy liquids are used for large pressure differences. For an indicating fluid 3 times heavier than water, the pressure measurement range is 3 times greater, but the resolution is reduced.

Indicating fluids can be colored water, oil, benzenes, bromides, and pure mercury. When selecting an indicating fluid, check the specifications for specific gravity, operating temperature range, vapor pressure, and flash point. Corrosive properties, solubility, and toxicity are also considerations.
Digital Manometers
A liquid manometer has limitations. Glass tubing, indicating fluids, and level mounting requirements are more suited to a laboratory than the field. Also, it cannot be interfaced with a computer or PLC. Such limitations can be overcome with digital manometers. These microprocessor-based instruments are available in convenient, portable sizes for ease of use in the field, or in panel or stand-alone mounting styles, with outputs for controlling a process or transferring measurement data.

Variations from standard conditions of density and gravity must be compensated for manually when making pressure measurements with liquid manometers. This is easier with digital manometers, because some of the correction factors for liquid manometers can be ignored and others can be compensated for in software.
With dual ports, swapping sensors is all that is needed to change among differential, gauge and absolute pressure measurements.
Other common features of digital manometers include:
  • Onboard memory for data logging or storing min./max. readings
  • Averaging a number of readings to dampen pressure pulses
Higher accuracy digital manometers are used to calibrate pressure transmitters and other pressure instrumentation in the field. Digital calibrators are faster and simpler as they require no boxes, gas cylinders, regulators, or weights to set up and have no special platforms or critical leveling requirements. Further comparisons of liquid and digital manometer specifications are shown in Table 1.
TABLE 1

Manometer Specifications


Liquid Manometers

Digital Manometers

U-tube

Well

Inclined

General Purpose

Calibrating

Range

100 in.

100 in.

20 in.

20-2000 in H2O,

20-2000 psig,

2000 mmHg

2000 in H2O,

2000 psig,

2000 mmHg

Accuracy

±½ of minor scale graduation

±½ of minor scale graduation

±½ of minor scale graduation

±0.025-0.1% F.S.

±0.025-0.1% F.S.

Wetted Parts
or Media
Compatibility

Cast iron, stainless steel, PVC, glass, Viton

Stainless steel, glass, Viton

Acrylic, stainless steel, aluminum, glass, Viton

Clean, dry non-corrosive gases; liquids compatible with stainless steel

Clean, dry non-corrosive gases; liquids compatible with stainless steel

Pressure
Rating

250 psig

250-500 psig

100-350 psig

2 × range

2 × range

Mounting

Wall, table

Wall, table, flush front, pipe

Wall, table

Portable

Portable

Relative Cost

Low

Low/medium

Medium

Medium

High
For Further Reading

Massey, B.S. 1989. Mechanics of Fluids, 6th Ed., London: Van Nostrand Reinhold.
Meriam Instrument. 1997. Using Manometers to Precisely Measure Pressure, Flow and Level, Cleveland: Meriam Instrument.
Meriam, J.B. 1938. The Manometer and Its Uses. 2nd Ed., Cleveland: Meriam Instrument.
Omega Engineering. 1999. Transactions in Measurement and Control: Force-Related Measurements, 2nd Ed. Stamford, CT: Putnam Publishing and Omega Press.
Yeager, John, and Hrusch-Tupta, M.A., Eds. 1998. Low Level Measurements. 5th Ed. Cleveland: Keithley Instruments.
SIDEBAR:
Manometer Pressure and Accuracy Glossary

Absolute Pressure. A measurement referenced to zero pressure; equals the sum of gauge pressure and atmospheric pressure. Common units are pounds per square inch (psia), millimeters mercury (mmHga), and inches mercury (in.Hga).
Accuracy. A measure of the closeness of agreement of a reading to that of a standard. For absolute accuracy, compare to a primary standard (one recognized by NIST). Accuracies are usually specified as a plus or minus percent of full scale. Calibration accuracies are often given as plus or minus percent of reading with plus or minus counts.
Ambient Pressure. The pressure of the medium surrounding a device. It varies from 29.92 in.Hg at sea level to a few inches at high altitudes.
Atmospheric Pressure. The pressure of the atmosphere on a unit surface. Also called barometric pressure. At sea level it is 29.92 in.Hg absolute.
Count. The smallest increment of an A/D conversion that is displayed.
Differential Pressure. The difference between two measurement points. Common units are inches of water (in.H2O), pounds per square inch (psi), and millibars (mbar).
Display Resolution. The maximum number of digits on a digital display. For example, a display resolution of 4½ digits reads a maximum of 19,999 counts; and a display resolution of 5 significant digits reads a maximum of 99,999 counts.
Gauge Pressure. A measurement referenced to atmospheric pressure. It varies with the barometric reading. Also used to specify the maximum pressure rating of manometers. Common units include pounds per square inch (psig).
Range. The region between the lower and upper limits of measurements.
Resolution. The smallest portion of a measurement that can be detected.
Sensitivity. The smallest change in measurement that can be detected.
Uncertainty. An estimate of the possible error in a measurement. This is the opposite of accuracy.
Vacuum. Any pressure below atmospheric pressure. When referenced to the atmosphere, it is called a vacuum (or negative gauge) measurement. When referenced to zero pressure, it is an absolute pressure measurement.
Zero Absolute Pressure. The complete absence of any gas; a perfect vacuum.